Gholographics Company

SPECTRO RADIOMETER

Gholographic Company can offers a fully automated, double monochromators base spectroradiometer GHPMS-700. It has high spectral resolution over a wide wavelength, and has excellent stray light control. These makes GHPMS-700 suitable for application with high requirements like photobiological safety analysis, and it has also very high accuracy when measuring general luminous flux and chromaticity parameters.

GHPMS-700 UV-VIS-nIR SPECTRORADIOMETER

Characteristics & Specifications:

  1. The Double-monochrometer & multi-grating & multi-photometer detector design makes low stray light noise and high measurement accuracy, and patent technology of SBCT enables wide dynamic range, it meets the CIE, IEC, ECE,EN and Chinese standards requirements.
  2. NIST(USA) and NIM (China) traceable.
  3. 16 bits high speed A/D, 32 bits software process.
  4. Wavelength: 200nm~800nm (special: 200nm~1400nm or 1400nm-3000nm).
  5. Accuracy of wavelength: 0.1nm (UV),0.2nm (VIS).
  6. Repeatability of wavelength:0.05nm (UV) 0.1nm (VIS).
  7. Spectrum sample interval: 0.1nm,1nm,5nm.
  8. Photometry linearity: 0.3%.
  9. Photometry repeatability: 0.1%.
  10. luminous flux range: 0.1lm~1.999x105lm ( Equipped with suitable integrating sphere).
  11. Accuracy of photometry: CLASS A.
  12. Accuracy of chromaticity coordinate: 0.0003 (Relative to NIM traceable illuminant A).
  13. Color temperature range: 1000K~25000K (Relative to NIM traceable illuminant A).
  14. Accuracy of color temperature: 0.3%.
  15. Rending index range: 0~100.0.
  16. Rending index error: (0.3%rd+0.5).
  17. Accuracy of color difference: 0.5 SDCM (Relative to NIM traceable illuminant A).
  18. Temperature range: 0 ~ 50.
  19. Temperature accuracy: 0.2.
  20. Stray light: 10-6.
  21. Photometry Dynamic range: 108.



Measurable items:

  1. Relative spectral radiant power P(¦Ë,or absolute spectral radiant power (W/nm), or absolute spectral irradiance (W/m2/nm),or absolute spectral radiance (W/m 2 /sr/nm).
  2. Chromaticity coordinate (x,y),(u,v), relative color temperature Tc, rending index Ra,Ri (i=1-15), color difference, peak wavelength, dominant wavelength,half width, color purity, red ratio.
  3. Temperature.
  4. Radiant power (W), irradiance (w/m 2 ), radiance(w/m 2/sr).
  5. Luminous flux (lm), illuminance (lx), luminance (cd/m 2).
  6. Irradiance
    • Irradiance UV (w/m2,200nm-400nm).
    • Irrradiance UV-A(w/m2,315nm-400nm).
    • Irradiance UV-B(w/m2,280nm-315nm).
    • Irradiance UV-C(w/m2,220nm-280nm).
    • Irradiance VIS(w/m2,380nm-800nm).
    • Irradiance blue¡±(w/m2,300nm-700nm).
  7. Radiance of blue¡±(w/m2/sr,300nm-700nm).

GHPMS-80 Sync-Skan SPECTRORADIOMETER

GHPMS-80 is a fast mechanical scanning spectrameter. It is the product of modern technologiesfor monochromators based spectroradiometer. GHPMS-80 has much faster speed than conventional monochromator based ones while the measurement accuracy is kept as high as originally. A typical measuring time during the spectral range from 380-800nm is only a few seconds comparing with several minutes of traditional scanning systems. Sync-Skan Speed the measurment largely.

Specifications:

  1. Sycn-skan technology is applied, and has faster test speed comparing with traditional scanning systems.
  2. A precise slit adjustment mechanism controlled by computer is set inside.
  3. Built-in wavelength calibration source.
  4. Spectrometer & Broadband-radiometer/photometer Combined Technique (SBCT) is adopted to enlarge the linear dynamic range.
  5. The calibration of spectral power distribution is directly traceable to NIST and NIM, superior accuracy can be achieved in UV-VIS-nIR measurement.
  6. Measure and display the relative spectral power distribution P(¦Ë), chromaticity coordinate(x , y)¡¢(u , v),correlative color temperature TC, rendering index Ra, Ri(i=1~15), color difference, peak wavelength, half width, color purity, dominant wavelength, red ratio, luminous flux etc.

Model Specifications Items
VIS
UV-VIS
VIS-nIR

Detector

PMT
PMT
Si

Monochromator

Wavelength range

380-800 nm
200-800nm
400-1100nm

Straylight(illuminant A)

1¡Á10-5
1¡Á10-5
1¡Á10-5

Bandpass

5/1/0.1nm
5/1/0.1nm
5/1/0.1nm

Sampling step

5/1/0.1nm
5/1/0.1nm
5/1/0.1nm

Wavelength accuracy

0.2nm
0.2nm
0.5nm

Wavelength reproducibility

0.1nm
0.1nm
0.2nm

Colortimetry parameters

Precision of Chromaticy (x,y)1)

0.0003
0.0004
0.0005

Tc range

1000-100000K

Tc accuracy

0.3%
0.4%
0.5%

Rendering index range

0 ~ 100.0

Rendering index accuracy

(0.3%rdg+0.5)

SDCM Accuracy

0.5SDCM
0.6SDCM
0.8SDCM

Photometry Parameters 2)

Photometry linearity

0.3%

Photometry range

Luminous flux:0.01lm-1.999¡Á105 lm (equipped with proper integrating sphere)

Photometer Head

Class A (standard Class in China)

Dynamic range

108

Electrical Data

AD converter

16 bits

PC Interface

RS-232-C

Miscellaneous

Power Input

"220V5%,50/60Hz (special order available)

Power Consumption

20VA(static), 50VA (dynamic)

Ambient Tempeerature

0-30¡æ, R.H 65%20%

Net Weight

Approx.30kg